Estimated Time To Complete

30 Hours

Accreditation

We are pleased to award category 1 credit(s) toward the AMA Physician's Recognition Award. By completing the Review in the CD-ROM Companion in Appendix XIV, you are eligible for up to 30 hours of category 1 credit. After answering all of the questions correctly, complete the review evaluation and enter the required identifying information on the certificate of course completion. This certificate is not valid until signed with authorized signature at the Foundation for Osteoporosis Research. The certificate may be printed one time only. Send the certificate and the required fee to the Foundation for Osteoporosis Research and Education for awarding of continuing education credits.

This activity is offered by the Foundation for Osteoporosis Research and Education, a CMA accredited provider. Physicians completing this course may report up to 30 hours of category 1 credits toward the California Medical Association's Certification in Continuing Education and the American Medical Association's Physician's Recognition award. Each physician should claim only those hours of credit that he/she actually spent in the activity.

This activity has been planned and implemented in accordance with the essential areas and policies of the Accreditation Council for Continuing Medical Education (ACCME) through the joint sponsorship of the Foundation for Osteoporosis Research and Education and The Humana Press. The Foundation for Osteoporosis Research and Education is accredited by the California Medical Association to provide continuing medical education for physicians.

Method of Participation

Read the book carefully. Complete the posttest and evaluation / certificate to be found on the companion CD-ROM. There is a $150 fee for this activity. Credit for the activity is available until September 15, 2006. Additional directions for obtaining credit can be found on the companion CD-ROM.

Faculty and Disclosure

Sydney Lou Bonnick, MD Institute for Women's Health Texas Woman's University Denton, Texas

Faculty for CME activities are expected to disclose to the activity audience any real or apparent conflict(s) of interest related to the content of the material they present. The following relationships have been disclosed:

Dr. Bonnick has nothing to disclose.

Provider Disclosure

The Foundation for Osteoporosis Research and Education is an independent organization that does not endorse specific products of any pharmaceutical concern and therefore has nothing to disclose. The Humana Press is an independent organization that does not endorse specific products of any pharmaceutical concern and therefore has nothing to disclose.

Intended Audience

This book is designed for physicians and technologists involved in the application of bone densitometry.

Overall Goal

The overall goal of this activity is to update the scientific knowledge and skills of physicians and technologists who utilize and interpret bone densitometry as part of patient management, in particular the management of patients who are at risk for or who have osteoporosis.

Learning Objectives

Upon completion of this continuing medical education activity, participants should have improved overall knowledge, skills, and attitudes concerning the use of bone densitometry. Specifically, the objectives are to:

1. Review the most clinically relevant aspects of interpreting bone density data.

2. Familiarize the physician with the resources found in the second edition of Bone Densitometry in Clinical Practice.

3. Emphasize the potential pitfalls in interpreting and reporting densitometry results.

4. Familiarize the physician with current recommendations and standards for patient selection for testing and for densitometry reporting.

Unlabeled/Unapproved Use Disclosure

In accordance with ACCME standards for Commercial Support, the audience is advised that this CME activity may contain references to unlabeled or unapproved uses of drugs or devices.

Densitometry Techniques

Contents

Plain Radiography in the Assessment of Bone Density Qualitative Morphometry Quantitative Morphometric Techniques Radiographic Photodensitometry Radiographic Absorptiometry Photon Absorptiometry Techniques Quantitative Ultrasound Bone Densitometry References

Clinical densitometry is relatively new but densitometry itself is actually quite old. It was first described more than 100 years ago in the field of dental radiology as dentists attempted to quantify the bone density in the mandible (1,2). With today's techniques, bone density can be quantified in almost every region of the skeleton. The extraordinary technical advances in recent years have expanded the realm of densitometry from that of a quantitative technique to that of an imaging technique as well. But even the oldest techniques remain both viable and valuable with computer modernization. Densitom-etry technologies have evolved as the understanding of relevant disease processes has increased. In a complimentary fashion, the understanding of the disease processes has increased as the technologies have evolved.

plain radiography in the assessment of bone density

The earliest attempts to quantify bone mineral density (BMD) utilized plain skeletal radiography. When viewed by the unaided eye, plain skeletal radiographs can only be used in an extremely limited fashion to quantify bone density. Demineralization becomes visually apparent only after 40% or more of the bone density has been lost (3). If demineralization is suspected from a plain film, a great deal of demineralization is presumed to have occurred. A more precise statement cannot be made. Plain radiographs have been used for qualitative and quantitative skeletal morphometry. plain radiographs were also used to assess bone density based on the optical densities of the skeleton when compared to simultaneously X-rayed standards of known density made from ivory or aluminum. With the advent of the photon absorptiometric techniques, most of these early methods as originally performed have fallen into disuse. Nevertheless, a brief review of these techniques should enhance the appreciation of the capabilities of modern testing and provide a background for understanding modern technologies.

Osteoporosis Trabecular Ray
Fig. 1-1. Quantitative spine morphometry. The vertebrae on this lateral lumbar spine X-ray demonstrate marked accentuation of the vertical trabecular pattern and thinning of the cortical shell. This is a Grade 2 spine.

qualitative morphometry

Qualitative Spinal Morphometry

Qualitative morphometric techniques for the assessment of bone density have been in limited use for more than 50 years. Grading systems for the spine relied on the appearance of the trabecular patterns within the vertebral body and the appearance and thickness of the cortical shell (4). Vertebrae were graded from IV down to I as the vertical trabecular pattern became more pronounced with the loss of the horizontal trabeculae and the cortical shell became progressively thinned. The spine shown in Fig. 1-1 demonstrates a pronounced vertical trabecular pattern. The cortical shell appears as though it was outlined in white around the more radiotranslucent vertebral body. These vertebrae would be classified as Grade II.

The Singh Index

The Singh Index is a qualitative morphometric technique that was similarly based on trabecular patterns, but based on those seen in the proximal femur (5). Singh and others noted that there was a predictable order in the disappearance of the five groups of trabeculae from the proximal femur in osteoporosis. Based on the order of disappearance, radiographs of the proximal femur could be graded 1 through 6 with lower values indicating a greater loss of the trabecular patterns normally seen in the proximal femur. Studies evaluating prevalent fractures demonstrated an association between Singh Index

Femur Ray Singh Index

Fig. 1-2. The Singh Index and calcar femorale thickness. A Grade 2 Singh Index would be assessed based on having only remnants of the principle compressive trabecular group visible. This is indicative of osteoporosis. The arrow points to the calcar femorale, which measured 4 mm in thickness. Values less than 5 mm are associated with hip fracture. This patient had experienced a contralateral hip fracture.

Fig. 1-2. The Singh Index and calcar femorale thickness. A Grade 2 Singh Index would be assessed based on having only remnants of the principle compressive trabecular group visible. This is indicative of osteoporosis. The arrow points to the calcar femorale, which measured 4 mm in thickness. Values less than 5 mm are associated with hip fracture. This patient had experienced a contralateral hip fracture.

values of 3 or less and the presence of fractures of the hip, spine, or wrist. Figure 1-2 shows a proximal femur with a Singh Index of 2. Only the trabecular pattern known as the principle compressive group, which extends from the medial cortex of the shaft to the upper portion of the head of the femur, remains. This patient was known to have osteoporotic spine fractures as well as a contralateral proximal femur fracture. Later attempts to demonstrate an association between Singh Index values and proximal femur bone density measured by dual-photon absorptiometry (DPA) were not successful (6).

Both of these qualitative morphometric techniques are highly subjective. In general, the best approach to their use required the creation of a set of reference radiographs of the various grades of vertebrae for spinal morphometry or proximal femurs for the Singh Index to which all other radiographs could be compared.

quantitative morphometric techniques

Calcar Femorale Thickness

A little known quantitative morphometric technique involved the measurement of the thickness of the calcar femorale. The calcar femorale is the band of cortical bone immediately above the lesser trochanter in the proximal femur. In normal subjects, this thickness is greater than 5 mm. In femoral fracture cases, it is generally less than 5 mm in thickness (7). The arrow, seen in Fig. 1-2 is pointing to the calcar femorale. This patient had previously suffered a femoral neck fracture. The thickness of the calcar femorale measured 4 mm.

Radiogrammetry

Radiogrammetry is the measurement of the dimensions of the bones using skeletal radiographs. Metacarpal radiogrammetry has been in use for almost 50 years. As originally practiced, the dimensions of the metacarpals were measured using a plain radiograph of the hand and fine calipers or a transparent ruler. The total width and medullary width of the metacarpals of the index, long, and ring fingers were measured at the midpoint of the metacarpal. The cortical width was calculated by subtracting the medullary width from the total width. Alternatively, the cortical width could be measured directly. A variety of different calculations were then made such as the metacarpal index (MI) and the hand score (HS). The MI is the cortical width divided by the total width. The HS, which is also known as the percent cortical thickness, is the MI expressed as a percentage. Measurements of the middle three metacarpals of both hands were also made and used to calculate the six metacarpal hand score (6HS). Other quantities derived from these measurements included the percent cortical area (%CA), the cortical area (CA), and the cortical area to surface area (CA/SA) ratio. The main limitation in all of these measurements is that they were based on the false assumption that the point at which these measurements were made on the metacarpal was a perfect hollow cylinder. Nevertheless, using these measurements and knowledge of the gravimetric density of bone, the bone density could be calculated. The correlation2 between such measurements and the weight of ashed bone was good, ranging from 0.79 to 0.85 (8,9). The precision of metacarpal radiogrammetry varied depending on the measurement used.2 The measurement of total width was very reproducible. The measurement of medullary width or the direct measurement of cortical width was less reproducible because the delineation between the cortical bone and medullary canal is not as distinct as the delineation between the cortical bone and soft tissue. Precision was variously reported as excellent to poor, but in expert hands it was possible to achieve a precision of 1.9% (10).

Although metacarpal radiogrammetry is an old technique and somewhat tedious to perform, it remains a viable means of assessing bone density in the metacarpals. Metacarpal radiogrammetry demonstrates a reasonably good correlation to bone density at other skeletal sites measured with photon absorptiometric techniques (11). The technique is very safe as the biologically significant radiation dose from a hand X ray is extremely low at only 1 mrem.

Radiogrammetry can also be performed at other sites such as the phalanx, distal radius, and femur (12-14). Combined measurements of the cortical widths of the distal radius and the second metacarpal are highly correlated with bone density in the spine as measured by DPA (12).

1 Correlation indicates the strength of the association between two values or variables. The correlation value is denoted with the letter r. A perfect correlation would be indicated by an r value of +1.00 or -1.00.

2 Techniques are compared on the basis of accuracy and precision, which can be described using the percent coefficient of variation (%CV). The %CV is the standard deviation (SD) divided by the average of replicate measurements expressed as a percentage. The lower the %CV, the better the accuracy or precision. See Chapters 3 and 11 for a detailed discussion of precision and accuracy.

Today, plain films of the hand and forearm can be digitized using flatbed optical scanners and radiogrammetry performed with computerized analysis of the digitized images. Using such a digital radiogrammetry (DXR) system, Bouxsein et al. (15) evaluated the utility of metacarpal radiogrammetry in predicting fracture risk and the correlation between metacarpal DXR-BMD and BMD measured by other techniques at other sites. The authors used a case-cohort approach to identify three groups of 200 women based on their having experienced a hip, wrist, or spine fracture during the first 5 years of the Study of Osteoporotic Fractures (16). DXR-BMD of the metacarpals was strongly correlated with distal and proximal radial BMD measured by single-photon absorptiometry (SPA)3 (r = 0.68 and 0.75, respectively). The correlation with femoral neck and lumbar spine BMD measured by dual energy X-ray absorptiometry (DXA)3 was more modest (r = 0.50 and 0.44, respectively). Metacarpal DXR-BMD predicted spine and wrist fracture risk as well as SPA BMD measurements of the distal or proximal radius or heel or DXA posterior-anterior (PA) lumbar spine or femoral neck. The increase in risk for wrist fracture was 1.6 for each SD decline in DXR-BMD and 1.9 for spine fracture. Although femoral neck BMD was the strongest predictor of hip fracture risk, metacarpal DXR-BMD predicted hip fracture risk as well as the other BMD measurements with an increase in risk of1.8 for each SD decline in BMD. This type of DXR system is available commercially from Sectra Pronosco in Denmark as part of a PACS4 system.

The Radiologic Osteoporosis Score

The radiologic osteoporosis score combined quantitative morphometry from three skeletal sites (14). Developed by Barnett and Nordin, this scoring system utilized radiogrammetry of the femoral shaft and metacarpal as well as an index of biconcavity of the lumbar vertebrae. In calculating what Barnett and Nordin called a peripheral score, the cortical thickness of the femoral shaft divided by the diameter of the shaft and expressed as a percentage was added to a similar measurement of the metacarpal. A score of 88 or less was considered to indicate peripheral osteoporosis. The biconcav-ity index was calculated by dividing the middle height of the third lumbar vertebra by its anterior height and expressing this value as a percentage. A biconcavity index of 80 or less indicated spinal osteoporosis. Combining both the peripheral score and bicon-cavity index resulted in the total radiologic osteoporosis score, which indicated osteoporosis if the value was 168 or less.

radiographic photodensitometry

Much of the development of the modern techniques of SPA, DPA and DXA actually came from early work on the X-ray-based method of photodensitometry (17). In photodensitometry, broad beam X-ray exposures of radiographs were obtained and the density of the skeletal image was quantified using a scanning photodensitometer. One such early device at Texas Woman's University is shown in Fig. 1-3. The effects of variations in technique such as exposure settings, beam energy, and film development were partially compensated by the simultaneous exposure of a step wedge of known densities on the film. An aluminum wedge was most often used, but other materials such as ivory were also employed (13). This technique could only be applied to areas of the

3 This technique is discussed later in this chapter.

4 Picture Archiving and Communications System.

Fig. 1-3. A radiographic photodensitometer at Texas Woman's University from the early 1950s.

skeleton in which the soft tissue coverage was less than 5 cm such as the hand, forearm, and heel. This restriction was necessary because of technical limitations from scattered radiation in thicker parts of the body and "beam hardening" or the preferential attenuation of the softer energies of the polychromatic X-ray beam as it passed through the body. Photodensitometry was also used in cadaver studies of the proximal femur (18). Such studies noted the predictive power for hip fracture of the density of the region in the proximal femur known as Ward's triangle5 30 years before studies using the modern technique of DXA in 1993 (19). The accuracy of such measurements was fairly good with a %CV of 5%. The correlation between metacarpal photodensitometry and ashed bone was also high at 0.88 (8). This was a slightly better correlation than seen with metacarpal radiogrammetry. The precision of photodensitometry was not as good, however, ranging from 5 to 15% (20). In this regard, the 6HS was superior (4). Radiation dose to the hand was the same for metacarpal radiogrammetry and radiographic photoden-sitometry. In both cases, the biologically significant radiation dose was negligible.

Radiographic photodensitometry was developed and used extensively by researchers Pauline Beery Mack and George Vose (21). Many of the original studies of the effects of weightlessness on the skeleton in the Gemini and Apollo astronauts were performed by Mack and her colleagues at Texas Woman's University (22). The photodensitometry hand film of one of the Gemini astronauts is shown in Fig. 1-4.

radiographic absorptiometry

Radiographic absorptiometry (RA) is the modern-day descendent of radiographic photodensitometry (23,24). The ability to digitize high resolution radiographic images

5 Ward's triangle was first described by F.O. Ward in Outlines of Human Osteology, London, Henry Renshaw, 1838. It is a triangular region created by the intersection of three groups of trabeculae in the femoral neck.

Radiographic Film Analysis

Fig. 1-3. A radiographic photodensitometer at Texas Woman's University from the early 1950s.

Metriscan Bone Density Result
Fig. 1-4. A radiographic photodensitometry hand film taken in 1965 of one of the Gemini astronauts. The Texas Woman's University aluminum wedge is seen next to the little finger.

and to perform computerized analysis of such images largely eliminated the errors introduced by differences in radiographic exposure techniques and overlying soft tissue thickness. In an early version of RA, two X-rays of the left hand using nonscreened film were taken at slightly different exposures. Standard X-ray equipment was used to perform the hand films. The initial recommended settings were 50 kVp at 300 mA for 1 second and 60 kVp at 300 mA for 1 second. The exact settings varied slightly with the equipment used and were adjusted so that the background optical density of each of the two hand films matched a quality control film. An aluminum alloy reference wedge was placed on the film prior to exposure, parallel to the middle phalanx of the index finger. After development, the films were sent to a central laboratory where they were digitized and analyzed by computer. The average BMD in arbitrary RA units of the middle phalanxes of the index, long, and ring fingers was reported. Figure 1-5 illustrates the X-ray appearance of the hand and aluminum alloy reference wedge.

In cadaveric studies, the accuracy of RA for the assessment of bone mineral content of the middle phalanges was good at 4.8% (25). The authors of this study noted that very thick soft tissue that might be seen in very obese subjects could potentially result in an underestimation of RA values. The correlation between the RA values and the ashed

Radiographic Absorptiometry
Fig. 1-5. A radiographic absorptiometry hand film. The small aluminum wedge, originally known as the Fel's wedge, is seen next to the index finger.

weight in the phalanges was excellent with r = 0.983. The short-term reproducibility of these measurements was also excellent at 0.6%.

The ability to predict bone density at other skeletal sites from hand RA is as good as that seen with other techniques such as SPA, DPA, DXA, or quantitative computed tomography (QCT) of the spine (23,26). This does not mean that RA hand values can be used to accurately predict bone density at other skeletal sites. Although the correlations between the different sites as measured by the various techniques are correctly said to be statistically significant, the correlations are too weak to allow clinically useful predictions of bone mass or density at one site from measurement at another.

The utility of modern-day RA in predicting hip fracture risk was suggested by an analysis of data acquired during the first National Health and Nutrition Examination Survey (NHANES I, 1971-1975). During this survey, 1559 hand radiographs of Caucasian women were obtained with the older technique of photodensitometry using the Texas Woman's University wedge (27). During a median follow-up of 14 years that extended through 1987, 51 hip fractures occurred. Based on radiographic photodensi-tometry of the second phalanx of the small finger of the left hand, the risk for hip fracture per SD decline in bone density increased 1.66-fold. These films were then re-analyzed using RA with some compensation for the differences in technique. This re-analysis yielded an increase in the risk for hip fracture per SD decline in RA bone density of 1.81fold. Huang and colleagues (28) evaluated the utility of RA in the prediction of vertebral fractures. They followed 560 postmenopausal women, average age 73.7 years, for an average of 2.7 years in the Hawaii Osteoporosis Study. The risk for vertebral fracture in this study using RA was 3.41-fold for each SD decline in bone density.

RA systems are commercially available. The automated Osteogram® system from Compumed, Inc. consists of the computer hardware, software, and film cassette with hand template and reference wedge needed to perform RA of the phalanges. A film-less, self-contained system is also in development. The Metriscan™ from Alara Inc. is a self-contained device that utilizes storage phosphor technology in place of X-ray film to perform RA of the phalanges. Both systems are discussed in Chapter 13.

photon absorptiometry techniques

In radiology, attenuation refers to a reduction in the number and energy of photons in an X-ray beam. Attenuation, then, is a reduction in an X-ray beam's intensity. To a large extent, the attenuation of X rays is determined by tissue density. The difference in tissue densities is responsible for creating the images seen on an X ray. The more dense the tissue, the more electrons it contains. The number of electrons in the tissue determines the ability of the tissue to either attenuate or transmit the photons in the X-ray beam. The differences in the pattern of transmitted or attenuated photons creates the contrast necessary to discern images on the X ray. If all the photons were attenuated (or none were transmitted), no image would be seen because the film would be totally white. If all of the photons were transmitted (or none were attenuated), no image would be seen because the film would be totally black. The difference in the attenuation of the X-ray photon energy by different tissues is responsible for the contrast on an X ray, which enables the images to be seen. If the degree of attenuation could be quantified, it would be possible to quantitatively assess the tissue density as well. This is the premise behind the measurement of bone density with photon absorptiometric techniques. The earliest photon absorptiometric techniques employed radionuclides to generate photon energy. These radionuclide-based techniques have given way to X-ray-based techniques. The basic principles on which they operate however, remain the same.

Single-Photon Absorptiometry

Writing in the journal Science in 1963, Cameron and Sorenson (29) described a new method for determining bone density in vivo by passing a monochromatic or single-energy photon beam through bone and soft tissue. The amount of mineral encountered by the beam could be quantified by subtracting the beam intensity after passage through the region of interest from the initial beam intensity. In these earliest SPA units, the results of multiple-scan passes at a single location, usually the midradius, were averaged (30). In later units, scan passes at equally spaced intervals along the bone were utilized such that the mass of mineral per unit of bone length could be calculated. A scintillation detector was used to quantify the photon energy after attenuation by the bone and soft tissue in the scan path. After the photon attenuation was quantified, a comparison to the photon attenuation seen with a calibration standard derived from dried defatted human ashed bone of known weight was made in order to determine the amount of bone mineral.

The photon beam and the detector were highly collimated or restricted in size and shape. The beam source and detector moved in tandem across the region of interest on the bone, coupled by a mechanical drive system. Iodine-125 at 27.3 keV, or americium-241 at 59.6 keV, were originally used to generate the single-energy photon beam although most SPA units subsequently developed in the United States employed only 125I.

Single Photon Absorptiometry Norland
Fig. 1-6. An early Norland model 2780 single photon absorptiometer. This device utilized 125I to generate photon energy. Photo courtesy of Norland, a CooperSurgical Company, Ft. Atkinson, WI.

The physical calculations for SPA determinations of bone mineral were valid only when there was uniform thickness of the soft tissue in the scan path. In order to artificially create this kind of uniform thickness, the limb to be studied had to be submerged in a water bath or surrounded by a tissue-equivalent material. As a practical matter, this limited SPA to measurements of the distal appendicular skeleton such as the radius and later, the calcaneus. Figure 1-6 is a photograph of an old SPA device, the Norland 2780, that was in use in the 1980s.

SPA was both accurate and precise, although the parameters varied slightly with the site studied. For SPA measurements of the midradius, accuracy ranged from 3 to 5% and precision, from 1 to 2% (29,31-33). Early measurements of the distal and ultradistal radius with SPA did not demonstrate the same high degree of precision primarily due to the marked changes in the composition of the bone with very small changes in location within the distal and ultradistal radius/ With later instruments that employed computer-enhanced localization routines and rectilinear scanning, SPA measurements of the distal and ultradistal radius approached a precision of 1% (34). Accuracy and precision of measurements at the calcaneus with SPA were reported to be less than 3% (32). The skin radiation dose for both the radius and calcaneus was 5 to 10 mrem (32,33). The biologically important radiation dose, the effective dose, was negligible. Results were reported as either bone mineral content (BMC) in grams or as BMC per unit length (BMD/l) in grams per centimeter. The time required to perform such studies was approximately 10 minutes (35).

SPA is rarely performed today, having been supplanted first by single-energy X-ray absorptiometry (SXA) and now DXA. The demise of SPA was due to improvements in ease of use and precision seen with SXA and DXA. SPA was an accurate technology that could be used to predict fracture risk. The ability to predict the risk of appendicular fractures with SPA measurements of the radius was convincingly established (36-38).

6 See Chapter 2 for a discussion of the composition of the radius and ulna.

SPA measurements of the radius were also good predictors of spine fracture risk and global7 fracture risk (36-40). Indeed, the longest fracture trials published to date, demonstrating the ability of a single bone mass measurement to predict fracture, were performed using SPA measurements of the radius.

Dual-Photon Absorptiometry

The basic principle involved in DPA for the measurement of bone density was the same as for SPA: quantifying the degree of attenuation of a photon energy beam after passage through bone and soft tissue. In DPA systems, however, an isotope that emitted photon energy at two distinct photoelectric peaks or two isotopes, each emitting photon energy at separate and distinct photoelectric peaks, were used. When the beam was passed through a region of the body containing both bone and soft tissue, attenuation of the photon beam occurred at both energy peaks. If one energy peak was preferentially attenuated by bone, however, the contributions of soft tissue to beam attenuation could be mathematically subtracted (41). As in SPA, the remaining contributions of beam attenuation from bone were quantified and then compared to standards created from ashed bone. The ability to separate bone from soft tissue in this manner finally allowed quantification of the bone density in those areas of the skeleton that were surrounded by large or irregular soft tissue masses, notably the spine and proximal femur. DPA was also used to determine total body bone density. The development of DPA and its application to the spine, proximal femur, and total body is attributed to a number of investigators: B.O. Roos, G.W. Reed, R.B. Mazess, C.R. Wilson, M. Madsen, W. Peppler, B.L. Riggs, W.L. Dunn, and H.W. Wahner (42-47).

The isotope most commonly employed in DPA in the United States was gadolinium-153, which naturally emitted photon energy at two photoelectric peaks, 44 keV and 100 keV. At the photoelectric peak of 44 keV, bone preferentially attenuated the photon energy. The attenuated photon beams were detected by a Nal scintillation detector and quantified after passage through pulse-height analyzers set at 44 and 100 keV. The shielded holder for the 153Gd source, which was collimated and equipped with a shutter that was operated by a computer, moved in tandem with the NaI detector in a rectilinear scan path over the region of interest. A point-by-point calculation of bone density in the scan path was made. Figure 1-7 is an intensity-modulated image of the spine created with an early DPA device.

DPA bone density studies of the lumbar spine were performed with the photon energy beam passing in a posterior to anterior direction. Because of the direction of the beam, the vertebral body and the posterior elements were included in the scan path. The transverse processes were eliminated. This resulted in a combined measurement of cortical and trabecular bone, or an integral measurement, that included the more trabecular vertebral body surrounded by its cortical shell and the highly cortical posterior elements. The results were reported as an areal density in g/cm2. The BMD of the proximal femur was also an areal density that was acquired with the beam passing in a posterior to anterior direction. Figure 1-8 shows an early DPA device with the patient positioned for a study of the lumbar spine.

7 Global fracture risk refers to the risk of having any and all types of fractures combined. This is in contrast to a site-specific fracture risk prediction in which the risk for a fracture at a specific skeletal site is given, such as spine fracture risk or hip fracture risk.

Hip Bone Density

Fig. 1-7. A dual-photon absorptiometry posterior-anterior spine study obtained on a device as shown in Fig. 1-8. The spine image is upside down. The histogram on the left was used to place the intervertebral disc space markers. The shortest bar in the vicinity of the disc space was identified and the marker was placed there.

Fig. 1-7. A dual-photon absorptiometry posterior-anterior spine study obtained on a device as shown in Fig. 1-8. The spine image is upside down. The histogram on the left was used to place the intervertebral disc space markers. The shortest bar in the vicinity of the disc space was identified and the marker was placed there.

Was this article helpful?

0 0
Essentials of Human Physiology

Essentials of Human Physiology

This ebook provides an introductory explanation of the workings of the human body, with an effort to draw connections between the body systems and explain their interdependencies. A framework for the book is homeostasis and how the body maintains balance within each system. This is intended as a first introduction to physiology for a college-level course.